Passive measurement of coarse particulate matter, PM 10 - 2.5

You are here

Date: 
Friday, February 1, 2008
Author(s): 
Darrin K. Ott
William Cyrs
Thomas M. Peters
Journal Title: 
Journal of Aerosol Science
Abstract: 

This work developed a passive sampling method to measure ambient PM10–2.5 . The coefficient of variation (CV) of PM10–2.5 measured with collocated passive samplers was 20.1% in laboratory tests and 11.6% in field tests. PM10–2.5 measured passively deviated from that measured with a filter-based dichotomous sampler by 29% in field tests. The strong correlation between PM10–2.5 derived passively with the dichotomous sampler (r=0.97 ) suggests that the bias observed in field tests was systematic and may be empirically corrected. The limit of detection of the method was determined to be   for a 5-day sample. A bootstrap analysis suggests that at least 300 particles should be imaged to stabilize the analytical variability of the method. Repeated analysis suggests that the CV for the analytical method is 5%. This method offers an inexpensive means to assess PM10–2.5 that may be incorporated into compliance networks or epidemiological studies.

Citation: 

Ott, Darrin K., William Cyrs, and Thomas M. Peters. "Passive measurement of coarse particulate matter, PM10–2.5." Journal of Aerosol Science 39, no. 2 (2008): 156-167. DOI: 10.1016/j.jaerosci.2007.11.002